15.Reference Level Accuracy, HP 8593EM, HP 8594EM, HP 8595EM, and HP 8596EM Table 2-16. Reference Level Accuracy, Linear Mode

Synthesizer/Level

EMC Analyzer

MKR 1 Reading (dB)

Generator Amplitude

Reference Level

 

 

 

 

 

 

 

 

(dBm)

(dB￿V)

Min.

TR Entry

Max.

 

 

 

 

 

010

87

0 (Ref)

0 (Ref)

0 (Ref)

0

 

 

 

 

97

00.4

10

+0.4

+10

107

00.5

11

+0.5

020

77

00.4

12

+0.4

030

67

00.5

13

+0.5

040

57

00.8

14

+0.8

050

47

01.0

15

+1.0

060

37

01.1

16

+1.1

070

27

01.2

17

+1.2

080

17

01.3

18

+1.3

 

 

 

 

 

Narrow Bandwidths

19. Press the following EMC analyzer keys:

4AMPLITUDE5 87 4+dB￿V5 SCALE LOG LIN (LOG) 1 4dB5

4BW5 IF BW AUTO MAN (MAN) 300 4Hz5

4SPAN5 10 4kHz5

4SWEEP/TRIG5 SWEEP CONT SGL (CONT)

20.Set the synthesizer/level generator to 010 dBm.

21.Set the 1 dB step attenuator to place the signal peak one to two dB (one to two divisions) below the reference level.

22.On the EMC analyzer, press the following keys:

4SGL SWP5

4MKR !5 MARKER ! HIGH

More 1 of 3 More 2 of 3 MARKER 1