57.Non-Harmonic Spurious Outputs, HP 8593EM, HP 8594EM, HP 8595EM, HP 8596EM Option 010

2.Connect the Type N cable between the RF OUT 50 ￿ and INPUT 50 ￿ connectors on the EMC analyzer. See Figure 2-82.

3.Press 4PRESET5 on the EMC analyzer, then wait for the preset routine to ￿nish. Set the EMC analyzer by pressing the following keys:

4FREQUENCY5 Band Lock 0-2.9 Gz BAND 0

The HP 8594EM does not need to be band locked.

4FREQUENCY5 300 4MHz5

4SPAN5 0 4Hz5

4BW5 IF BW AUTO MAN (MAN) 30 4kHz5

4MKR5

4AUX/USER5 Track Gen SRC PWR ON OFF (ON) 102 4dB￿V5 TRACKING PEAK

Wait for the PEAKING message to disappear, then press the following keys:

SRC PWR ON OFF (ON) 106 4dB￿V5

4SGL SWP5

4.Connect the Type N cable from the tracking generator output to the microwave spectrum analyzer RF INPUT as shown in Figure 2-82.

2-362 Performance Veri￿cation Tests