20. Frequency Response, HP 8593EM

20.Frequency Response , HP 8593EM

This performance veri￿cation test must be performed with the EMC analyzer set in the spectrum analyzer mode.

The output of the synthesized sweeper is fed through a power splitter to a power sensor and the EMC analyzer. The synthesized sweeper power level is adjusted at 300 MHz to place the displayed signal at the analyzer center horizontal graticule line. The measuring receiver, used as a power meter, is placed in RATIO mode. At each new synthesized sweeper frequency and analyzer center frequency setting, the synthesized sweeper power level is adjusted to place the signal at the center horizontal graticule line. The measuring receiver displays the inverse of the frequency response relative to 300 MHz (CAL OUT frequency).

Equipment Required

Synthesized sweeper

Measuring receiver (used as a power meter)

Synthesizer/Level generator

Power sensor, 50 MHz to 26.5 GHz

Power splitter

Termination, 50 ￿

Adapter, Type N (m) to APC 3.5 (m)

Adapter, Type N (f) to BNC (f)

Adapter, 3.5 mm (f) to 3.5mm (f)

Adapter, Type BNC (f) to SMA (m)

Cable, BNC, 122 cm (48 in)

Cable, APC 3.5, 91 cm (36 in)

Performance Veri￿cation Tests 2-105