7.System Related Sidebands, HP 8590 EM-Series

8.Set the EMC analyzer to measure the system related sideband below the signal by pressing the following EMC analyzer keys:

4 + 5 (step-down key)

4 + 5 (step-down key)

9.Measure the system related sideband below the signal by pressing 4SGL SWP5. Wait for the completion of a new sweep, then press 4MKR !5 MKR ! HIGH .

Record the Marker-1 Amplitude as TR Entry 2 of the performance veri￿cation test record.

The system related sideband below the signal should be > 65 dBc.

Performance Veri￿cation Tests 2-31