55. Harmonic Spurious Outputs, HP 8593EM,
HP 8594EM, HP 8595EM, and HP 8596EM Option 010
55.Harmonic Spurious Outputs , HP 8593EM, HP 8594EM,HP 8595EM, and HP 8596EM Option 010This performance verication test must be performed with the EMC analyzer set in the spectrum analyzer mode.
The tracking generator output is connected to the EMC analyzer input and the tracking is adjusted at 300 MHz for a maximum signal level. The tracking generator output is then connected to the input of a microwave spectrum analyzer. The tracking generator is tuned to several dierent frequencies and the amplitude of the second and third harmonics relative to the fundamental are measured at each frequency.
There are no related adjustment procedures for this performance verication test.
Equipment Required
Spectrum analyzer, microwave
Cable, Type N, 62 cm (24 in)
Cable, BNC, 23 cm (9 in)
Adapter, Type N (m) to BNC (f)
Procedure
Note | It is only necessary to perform Step 1 if more than two hours |
| have elapsed since a |
| spectrum analyzer was performed. |
| The microwave spectrum analyzer should be allowed to warm |
| up for at least 30 minutes before proceeding. |
|
|
1.Perform a
Note that the following steps are for an HP 8566A/B microwave spectrum analyzer, the steps may be dierent if you are using another microwave spectrum analyzer.
a. Connect a BNC cable between the CAL OUTPUT and the RF INPUT.