62. CISPR Pulse Response, HP 8590 EM-Series

4AMPLITUDE5 REF LVL 67 4dB￿V5

27.Set the PERIOD to 10 ms on the pulse generator.

28.Press the following EMC analyzer keys:

4SGL SWP5

MARKER ! HIGH

29.Record the marker amplitude reading in Table 2-82 as the Measured Relative Equivalent Level of Pulse for Band A, 100 Hz Repetition Frequency.

30.Set the PERIOD to 16.7 ms on the pulse generator.

31.Press the following EMC analyzer keys:

4SGL SWP5

4MKR !5 MARKER ! HIGH

32.Record the marker amplitude reading in Table 2-82 as the Measured Relative Equivalent Level of Pulse for Band A, 60 Hz Repetition Frequency.

33.Set the PERIOD to 100 ms on the pulse generator.

34.Press the following EMC analyzer keys:

4SGL SWP5

MARKER ! HIGH

35.Record the marker amplitude reading in Table 2-82 as the Measured Relative Equivalent Level of Pulse for Band A, 10 Hz Repetition Frequency.

36.Set the PERIOD to 200 ms on the pulse generator.

37.Press the following EMC analyzer keys:

4SGL SWP5

MARKER ! HIGH

38.Record the marker amplitude reading in Table 2-82 as the Measured Relative Equivalent Level of Pulse for Band A, 5 Hz Repetition Frequency.

39.Set the PERIOD to 500 ms on the pulse generator.

40.Press the following EMC analyzer keys:

Performance Veri￿cation Tests 2-391