32. Spurious Response, HP 8595EM

16.Press 4MKR !5 MARKER ! HIGH , More 1 of 3 More 2 of 3 MARKER 1 , then record the power meter reading at 2.8 GHz in Table 2-48.

17.Set the synthesized sweeper CW to 5.6 GHz.

18.Press the following EMC analyzer keys:

4FREQUENCY5 5.6 4GHz5

4MKR !5 MARKER ! HIGH

4AMPLITUDE5 More 1 of 3 PRESEL PEAK Wait for the peaking message to disappear.

4MKR !5 MARKER ! HIGH

4MKR5 More 1 of 3 MK TRACK ON OFF (ON)

19.Adjust the synthesized sweeper power level until the Marker 1 Amplitude reads 0 dB 60.20 dB.

20.Enter the power sensor 6 GHz Cal Factor into the power meter.

21.Record the Power Meter Reading at 5.6 GHz in Table 2-48.

22.Subtract the Power Meter Reading at 5.6 GHz from the Power Meter Reading at 2.8 GHz, then record this value as the Frequency Response Error (FRE) in Table 2-48.

For example, if the Power Meter Reading at 5.6 GHz is 06.45 dBm and the Power Meter Reading at 2.8 GHz is 07.05 dBm, the Frequency Response Error would be 07.05 dBm 0 (06.45 dBm) = 00.60 dB.

Power Meter Reading at 2.8 GHz 0 Power Meter Reading at 5.6 GHz =

FRE

Performance Veri￿cation Tests 2-211