32. Spurious Response, HP 8595EM
16.Press 4MKR !5 MARKER ! HIGH , More 1 of 3 More 2 of 3 MARKER 1 , then record the power meter reading at 2.8 GHz in Table
17.Set the synthesized sweeper CW to 5.6 GHz.
18.Press the following EMC analyzer keys:
4FREQUENCY5 5.6 4GHz5
4MKR !5 MARKER ! HIGH
4AMPLITUDE5 More 1 of 3 PRESEL PEAK Wait for the peaking message to disappear.
4MKR !5 MARKER ! HIGH
4MKR5 More 1 of 3 MK TRACK ON OFF (ON)
19.Adjust the synthesized sweeper power level until the Marker 1 Amplitude reads 0 dB 60.20 dB.
20.Enter the power sensor 6 GHz Cal Factor into the power meter.
21.Record the Power Meter Reading at 5.6 GHz in Table
22.Subtract the Power Meter Reading at 5.6 GHz from the Power Meter Reading at 2.8 GHz, then record this value as the Frequency Response Error (FRE) in Table
For example, if the Power Meter Reading at 5.6 GHz is 06.45 dBm and the Power Meter Reading at 2.8 GHz is 07.05 dBm, the Frequency Response Error would be 07.05 dBm 0 (06.45 dBm) = 00.60 dB.
Power Meter Reading at 2.8 GHz 0 Power Meter Reading at 5.6 GHz =
FRE
Performance Verication Tests