15. Reference Level Accuracy, HP 8593EM, HP 8594EM, HP 8595EM, and HP 8596EM
4.Set the 1 dB step attenuator to place the signal peak one to two dB (one to two divisions) below the reference level.
5.On the EMC analyzer, press the following keys:
4SGL SWP5
4MKR !5 MARKER ! HIGH
More 1 of 3 More 2 of 3 MARKER 1
6.Set the synthesizer/level generator amplitude and EMC analyzer reference level according to Table
7.At each setting, press 4SGL SWP5 on the EMC analyzer.
8.Record the MKR 1 amplitude reading in the performance verication test record as indicated in Table
9.Repeat steps 6 through 8 for each entry in Table
Table
Synthesizer/Level | EMC Analyzer | MKR 1 Reading (dB) | ||
Generator Amplitude | Reference Level |
|
|
|
|
|
|
|
|
(dBm) | (dBV) | Min. | TR Entry | Max. |
|
|
|
|
|
010 | 87 | 0 (Ref) | 0 (Ref) | 0 (Ref) |
0 | 97 | 00.4 | 1 | +0.4 |
+10 | 107 | 00.5 | 2 | +0.5 |
020 | 77 | 00.4 | 3 | +0.4 |
030 | 67 | 00.5 | 4 | +0.5 |
040 | 57 | 00.8 | 5 | +0.8 |
050 | 47 | 01.0 | 6 | +1.0 |
060 | 37 | 01.1 | 7 | +1.1 |
070 | 27 | 01.2 | 8 | +1.2 |
080 |
|
|
|
|
17 | 01.3 | 9 | +1.3 | |
|
|
|
|
|