25. Other Input Related Spurious Responses, HP 8593EM
Figure
Other Input Related Spurious Responses Test Setup, HP 8593EM
Procedure
Band 0
1.Zero and calibrate the measuring receiver and power sensor in log mode (power reads out in dBm), as described in the measuring receiver operation manual. Enter the power sensor 2 GHz Cal Factor into the measuring receiver.
2.Press INSTRUMENT PRESET on the synthesized sweeper and set the controls as follows:
CW :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: :: :: 2000 MHz
POWER LEVEL : ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: 04 dBm
3.Connect the equipment as shown in Figure
Option 026 only: Connect the power splitter to the EMC analyzer input directly. Do not use an adapter to connect the power splitter to the EMC analyzer.
4.Press 4PRESET5 on the EMC analyzer, then wait for the preset routine to nish. Set the EMC analyzer by pressing the following keys:
4FREQUENCY5 2.0 4GHz5