62. CISPR Pulse Response, HP 8590
Isolation Check
Figure
11.Connect the equipment as shown in Figure
12.On the EMC analyzer, press the following keys:
4MODE5 EMC ANALYZER
4PRESET5 (Wait for the preset routine to nish.)
4FREQUENCY5 CENTER FREQ 50 4MHz5
4SPAN5 1 4MHz5
4AMPLITUDE5 SCALE LOG LIN (LIN)
More 1 of 3 Amptd Units dBV
4AMPLITUDE5 SCALE LOG LIN (LOG)
More 1 of 3 Amptd Units dBV
4MKR !5 MARKER ! HIGH
4SAVE/RECALL5 Save Internal STATE ! INTRNL 1 4MKR !5 More 1 of 3 MARKER ! REF LVL 4MKR5 MARKER 1