This performance verication test must be performed with the EMC analyzer set in the spectrum analyzer mode.
The EMC analyzer input is terminated and the EMC analyzer is swept from 150 kHz to 1 MHz. Then the EMC analyzer is swept in 10 MHz spans throughout the 1 MHz to 1.8 GHz range. Any responses above the specication are noted.
There are no related adjustment procedures for this performance test.
Equipment
Termination, 50
Figure
Procedure
150 kHz to 1 MHz
1.Connect the termination to the EMC analyzer input as shown in Figure
2.Press 4PRESET5 on the EMC analyzer, then wait for the preset routine to nish. Press the following EMC analyzer keys:
4FREQUENCY5
START FREQ 150 4kHz5
STOP FREQ 1 4MHz5
4AMPLITUDE5 47 4dBV5 ATTEN AUTO MAN (MAN) 0 4dB5
4BW5 300 4Hz5
4DISPLAY5 DSP LINE ON OFF (ON) 17 4dBV5