17.Resolution (IF) Bandwidth Accuracy, HP 8590
25.Using the synthesizer/level generator knob, increase the frequency so that the
26.Record the synthesizer/level generator frequency readout in column 2 of Table
27.Adjust the synthesizer/level generator frequency for maximum marker amplitude.
28.Repeat steps 18 through 26 for the 120 kHz EMI bandwidth (4BW5
120kHz EMI BW ) and the 1 MHz EMI bandwidth (4BW5 1 4MHz5).
29.Subtract the Synthesizer Lower Frequency from the Synthesizer Upper Frequency. Record the dierence as the Resolution Bandwidth Accuracy, in the performance verication test record as indicated in Table
RES BW Accuracy = Upper Frequency 0 Lower Frequency
Table
EMC Analyzer |
| Column 1 |
| Column 2 | TR Entry | ||
RES BW |
| Synthesizer |
| Synthesizer |
| ||
|
| Lower Frequency |
| Upper Frequency |
| ||
|
|
|
|
|
|
|
|
9 kHz |
|
|
|
|
|
| 8 |
120 kHz |
|
|
|
|
|
| 9 |
1 MHz |
|
|
|
|
|
| 10 |
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
30.Press 4PRESET5 on the EMC analyzer, then wait for the preset routine to nish. Set the EMC analyzer by pressing the following keys:
4FREQUENCY5 50 4MHz5
4SPAN5 1 4MHz5
4MKR !5 MARKER ! HIGH
4MKR5 More 1 of 3 MK TRACK ON OFF (ON)
4SPAN5 1 4kHz5
Wait for the auto zoom routine to nish, then press the following keys:
4MKR5 MARKER 1 ON OFF (OFF)
Performance Verication Tests