18. Calibrator Amplitude Accuracy, HP 8590
Figure
11.Connect the equipment as shown in Figure
12.On the power meter, press the dBm mode key. Record the Power Meter Reading in dBm in the worksheet as the Power Meter Reading.
13.Subtract the Corrected Insertion Loss (step 10) from the Power Meter Reading (step 12).
CAL OUT Power = Power Meter Reading 0 Corrected Insertion Loss
Example: If the Corrected Insertion Loss is 010.0 dB, and the power meter reading is 030 dB, then (030 dB) 0 (010.0 dB) = 020 dB
14.Record this value as TR Entry 1 of the performance verication test record as the CAL OUT power. The CAL OUT should be 020 dBm 60.4 dB.