62. CISPR Pulse Response, HP 8590
CW Measurement for 9 kHz EMI Bandwidth
50.Press 4PRESET5 on the EMC analyzer. Wait for the preset routine to nish.
51.Press 4RECALL5 1 on the pulse generator.
52.Press 4RECALL5 2 on the synthesizer/level generator.
53.Press the following keys on the EMC analyzer:
4SAVE/RECALL5 Recall Internal INTERNAL ! STATE 1
4MKR5 MARKER NORMAL
4SPAN5 20 4kHz5
4BW5 9 kHz EMI BW
4DET5 QP ON OFF (ON)
Note that this routine takes approximately 1 minute to execute.
4AMPLITUDE5 REF LVL 67 dBV
4MKR !5 MARKER ! HIGH
54.Record the
9 kHz Pulse RF Signal Setup
55.Press 4RECALL5 1 on the pulse generator. Set the pulse generator to the following conditions. (Use the 4CHS5 key to change the sign of the value entered on the pulse generator.)
PER :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: : | 10 ms |
WID ::: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: : | 2.2 s |
LOL :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: : | 01.7 V |
56.Press 4RECALL5 1 on the synthesizer/level generator. Set the synthesizer/level generator amplitude to the required amplitude value for the 9 kHz lter recorded in Table