62. CISPR Pulse Response, HP 8590 EM-Series

4DET5 QP/AVG 10X OFF (10X)

4SGL SWP5

4MKR !5 MARKER ! HIGH

41.Record the marker amplitude reading in Table 2-82 as the Measured Relative Equivalent Level of Pulse for Band A, 2 Hz Repetition Frequency.

42.Set the PERIOD to 980 ms on the pulse generator.

43.Press the following EMC analyzer keys:

4SGL SWP5

MARKER ! HIGH

44.Record the marker amplitude reading in Table 2-82 as the Measured Relative Equivalent Level of Pulse for Band A, 1 Hz Repetition Frequency.

45.Press 4TRIG5 on the pulse generator.

46.Press 4SGL SWP5 on the EMC analyzer.

47.Let the EMC analyzer sweep 3 divisions then press 4MAN5 on the pulse generator.

48.Press MARKER ! HIGH .

49.Record the Marker reading for Isolated Pulse Measurement for Band A in Table 2-82.

2-392 Performance Veri￿cation Tests