14. Reference Level Accuracy, HP 8591EM
18.Record the MKR 1 amplitude reading in the performance verication test record as indicated in Table
Table
Reference Level Accuracy,
Log Mode for Narrow IF Bandwidths
Synthesizer/Level | EMC Analyzer | MKR 1 Reading (dB) | ||
Generator Amplitude | Reference Level |
|
|
|
|
|
|
|
|
(dBm) | (dBV) | Min. | TR Entry | Max. |
|
|
|
|
|
010 | 87 | 0 (Ref) | 0 (Ref) | 0 (Ref) |
0 | 97 | 00.4 | 19 | +0.4 |
+10 | 107 | 00.5 | 20 | +0.5 |
020 | 77 | 00.4 | 21 | +0.4 |
030 | 67 | 00.5 | 22 | +0.5 |
040 | 57 | 00.8 | 23 | +0.8 |
050 | 47 | 01.1 | 24 | +1.1 |
060 | 37 | 01.2 | 25 | +1.2 |
070 | 27 | 01.3 | 26 | +1.3 |
080 | 17 | 01.4 | 27 | +1.4 |
|
|
|
|
|
19.Repeat steps 8 through 13, using Table
20.Record the MKR 1 amplitude reading in the performance verication test record as indicated in Table