44. Residual Responses, HP 8591EM
10.Reconnect the termination to the EMC analyzer input as shown in Figure
11.Press the following EMC analyzer keys:
4FREQUENCY5 5 4MHz5
4AMPLITUDE5 47 4dBV5
4SWEEP/TRIG5 SWEEP CONT SGL (CONT)
44. Residual Responses, HP 8591EM
10.Reconnect the termination to the EMC analyzer input as shown in Figure
11.Press the following EMC analyzer keys:
4FREQUENCY5 5 4MHz5
4AMPLITUDE5 47 4dBV5
4SWEEP/TRIG5 SWEEP CONT SGL (CONT)