50. Absolute Amplitude Accuracy, HP 8593EM, HP 8594EM, HP 8595EM, HP 8596EM Option 010
7.Record the power level displayed on the measuring receiver as the Absolute Amplitude Accuracy in the performance verication test record as
TR Entry 1.
8.Press RATIO on the measuring receiver. Power levels will now readout in dB relative to the power level just measured at the 87 dBV (020 dBm) output power level setting.
9.Set the EMC analyzer SRC POWER to the settings indicated in Table
10.Calculate the Absolute Vernier Accuracy by subtracting the SRC POWER setting (in dBm) from the Measured Power Level for each SRC POWER setting in Table
Measured Power Level 0 SRC POWER 0 20 = Absolute Vernier Accuracy
For example: Where the SRC POWER = 021 dBm: 00.9 0 (021) 0 20 = 0.1
11.Calculate the
Start by subtracting the Absolute Vernier Accuracy for the 017 dBm (90 dBV) SRC POWER setting from the Absolute Vernier Accuracy for the 018 dBm (89 dBV) setting.
Record this calculation in the
12.Locate the most positive Absolute Vernier Accuracy value in Table
13.Locate the most negative Absolute Vernier Accuracy value in Table
14.Locate the largest
15.Locate the smallest