32. Spurious Response, HP 8595EM
52.Disconnect the power sensor from the directional coupler. Connect the directional coupler directly to the EMC analyzer INPUT 50 using an adapter (do not use a cable).
53.On the EMC analyzer, press the following key:
4MKR !5 MARKER ! HIGH
4AMPLITUDE5 More 1 of 3 PRESEL PEAK Wait for the peaking message to disappear.
4MKR5 More 1 of 3 MK TRACK ON OFF (ON)
4SPAN5 200 4kHz5
Wait for the AUTO ZOOM message to disappear, then press the following EMC analyzer keys:
4MKR5 More 1 of 3 MK TRACK ON OFF (OFF) 4FREQUENCY5 4 * 5
4MKR !5 MARKER ! HIGH
More 1 of 3 MARKER ! REF LVL
54.On synthesized sweeper #2, set RF on. Adjust the power level until the two signals are displayed at the same amplitude.
If necessary, adjust the EMC analyzer center frequency until the two signals are centered on the display.
55.Set the EMC analyzer by pressing the following keys:
4BW5 IF BW AUTO MAN (MAN) 1 4kHz5
AVG BW AUTO MAN (MAN) 100 4Hz5
56.Press 4MKR !5 MARKER ! HIGH , More 1 of 3 More 2 of 3 MARKER 1 then set the DISPLAY LINE to a value 54 dB below the current reference level setting.
The