15. Reference Level Accuracy, HP 8593EM, HP 8594EM, HP 8595EM, and HP 8596EM
Linear Scale
10.Set the synthesizer/level generator amplitude to 010 dBm.
11.Set the 1 dB step attenuator to 0 dB attenuation.
12.Set the EMC analyzer controls as follows:
4AMPLITUDE5 87 4+dBV5
SCALE LOG LIN (LIN)
4AMPLITUDE5 More 1 of 3 Amptd Units 4+dBV5 4SWEEP/TRIG5 SWEEP CONT SGL (CONT)
4MKR5 More 1 of 3 MARKER ALL OFF
13.Set the 1 dB step attenuator to place the signal peak one to two divisions below the reference level.
14.On the EMC analyzer, press the following keys:
4SGL SWP5
4MKR !5 MARKER ! HIGH
More 1 of 3 More 2 of 3 MARKER 1
15.Set the synthesizer/level generator amplitude and EMC analyzer reference level according to Table
16.At each setting, press 4SGL SWP5 on the EMC analyzer.
17.Record the MKR 1 amplitude reading in the performance verication test record as indicated in Table
18.Repeat steps 15 through 17 for each entry in Table
Performance Verication Tests