54. Harmonic Spurious Outputs, HP 8591EM Option 010
Procedure
1.Connect the Type N cable between the RF OUT 50 and INPUT 50 connectors on the EMC analyzer. See Figure
2.Press 4PRESET5 on the EMC analyzer, then wait for the preset routine to nish. Set the EMC analyzer by pressing the following keys:
4FREQUENCY5 300 4MHz5
4SPAN5 ZERO SPAN
4MKR5
4AUX/USER5 Track Gen SRC PWR ON OFF (ON) 102 4dBV5
3.On the EMC analyzer, press TRACKING PEAK . Wait for the PEAKING message to disappear, then press the following keys:
107 4dBV5
4FREQUENCY5 10 4MHz5
4SGL SWP5
It is only necessary to perform the next step if more than two hours have elapsed since a
The microwave spectrum analyzer should be allowed to warm up for at least 30 minutes before proceeding.
4.Perform a
Note that the following steps are for an HP 8566A/B microwave spectrum analyzer, the steps may be dierent if you are using another microwave spectrum analyzer.
a.Connect a BNC cable between the CAL OUTPUT and the RF INPUT.
b.Press 42 0 22 GHz5 (INSTR PRESET), 4RECALL5, 8. Adjust AMPTD CAL for a marker amplitude reading of 010 dBm.
c.Press 4RECALL5, 9. Adjust FREQ ZERO for a maximum amplitude response.
5.Connect the Type N cable from the tracking generator output to the microwave spectrum analyzer RF INPUT as shown in Figure
6.Set the microwave spectrum analyzer controls as follows:
CENTER FREQUENCY : :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: : : ::: :: :: :: : 10 MHz
SPAN ::: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: : 100 kHz
Performance Verication Tests