18. Calibrator Amplitude Accuracy, HP 8590 EM-Series

Part 1: LPF, Attenuator and A dapter Insertion Loss Characterization

1.Zero and calibrate the measuring receiver and 100 kHz to 1800 MHz power sensor in LOG mode as described in the measuring receiver operation manual.

CAUTION

Do not attempt to calibrate the low-power power sensor without

 

the reference attenuator or damage to the low-power power

 

sensor will occur.

 

 

2.Zero and calibrate the power meter and low-power power sensor, as described in the power meter operation manual.

3.Press INSTRUMENT PRESET on the synthesized sweeper, then set the controls as follows:

CW : : : : :: :: : :: : : : : : : : :: : : : : : : : : : :: :: : : : :: : : : : : : : : : : : : : : : : : : :300 MHz

POWER LEVEL :: : : : : : : : : : : : : : : : : : : : : : : :: : : : : : : : :: : : : :: : : : 015 dBm

Figure 2-21. LPF Characterization

4.Connect the equipment as shown in Figure 2-21. Connect the low-power power sensor directly to the power splitter (bypass the LPF, attenuator, and

2-94 Performance Veri￿cation Tests