19. Frequency Response, HP 8591EM

The result should be less than 2.0 dB.

The absolute values in steps 3 and 6 should be less than 1.5 dB.

Table 2-24. Frequency Response Errors Worksheet

Column 1

 

Column 2

Column 3

 

Column 1

 

Column 2

Column 3

EMC Analyzer

Error Relative

CAL FACTOR

 

EMC Analyzer

Error Relative

CAL FACTOR

Frequency

 

to 300 MHz

Frequency

 

Frequency

 

to 300 MHz

Frequency

(MHz)

 

(dB)

(GHz)

 

(MHz)

 

(dB)

(GHz)

 

 

 

 

 

 

 

 

 

 

 

50

 

 

 

0.03

 

950

 

 

 

1.0

100

 

 

 

0.1

 

1000

 

 

 

1.0

150

 

 

 

0.1

 

1050

 

 

 

1.0

200

 

 

 

0.3

 

1100

 

 

 

1.0

250

 

 

 

0.3

 

1150

 

 

 

1.0

300 (Ref)

 

 

 

0.3

 

1200

 

 

 

1.0

350

 

 

 

0.3

 

1250

 

 

 

1.0

400

 

 

 

0.3

 

1300

 

 

 

1.0

450

 

 

 

0.3

 

1350

 

 

 

1.0

 

 

 

 

 

 

500

 

 

 

 

 

1400

 

 

 

1.0

 

 

 

0.3

 

 

 

 

550

 

 

 

1.0

 

1450

 

 

 

1.0

600

 

 

 

1.0

 

1500

 

 

 

1.0

650

 

 

 

1.0

 

1550

 

 

 

2.0

700

 

 

 

1.0

 

1600

 

 

 

2.0

750

 

 

 

1.0

 

1650

 

 

 

2.0

800

 

 

 

1.0

 

1700

 

 

 

2.0

850

 

 

 

1.0

 

1750

 

 

 

2.0

900

 

 

 

1.0

 

1800

 

 

 

2.0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Performance Veri￿cation Tests 2-103