20. Frequency Response, HP 8593EM
Frequency Response , Band 3
1.Enter the most positive number from Table
2.Enter the most negative number from Table
3.Subtract step 2 from step 1. Enter this value as TR Entry 12 of the performance verication test record.
Frequency Response , Band 4
Option 026 or 027 only: Proceed to \Frequency Response, Band 4 for Option 026 or 027" if the EMC analyzer is equipped with Option 026 or 027.
1.Enter the most positive number from Table
2.Enter the most negative number from Table
3.Subtract step 2 from step 1. Enter this value as TR Entry 15 of the performance verication test record.
Frequency Response, Band 4 for Option 026 or 027
1.Enter the most positive number from Table
2.Enter the most negative number from Table
3.Subtract step 2 from step 1. Enter this value as TR Entry 15 of the performance verication test record.
Performance Verication Tests