26.Other Input Related Spurious Responses, HP 8594EM
26.Other Input Related Spurious Responses ,HP 8594EMThis performance verication test must be performed with the EMC analyzer set in the spectrum analyzer mode.
A synthesized source and the EMC analyzer are set to the same frequency and the amplitude of the source is set to 0 dBm (107 dBV). A marker amplitude reference is set on the analyzer. The source is then tuned to several dierent frequencies which should generate image, multiple, and
There are no related adjustment procedures for this performance verication test.
Equipment Required
Synthesized sweeper
Measuring receiver (used as a power meter)
Power sensor, 50 MHz to 2.9 GHz
Power splitter
Adapter, Type N (m) to APC 3.5 (m)
Adapter, APC 3.5 (f) to APC 3.5 (f)
Cable, APC 3.5, 91 cm (36 in)
Performance Verication Tests