8.Frequency Span Readout Accuracy, HP 8591EM
3.Press INSTRUMENT PRESET on the synthesized sweeper, then set the controls as follows:
CW :: :: :: ::: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: 74 MHz
POWER LEVEL :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: 05 dBm
4. Set the synthesizer/level generator controls as follows:
FREQUENCY :: :: :: :: ::: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: : 66 MHz
AMPLITUDE : :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: : 0 dBm
5.Adjust the EMC analyzer center frequency to center the two signals on the display.
6.On the EMC analyzer, press 4SGL SWP5. Wait for the completion of a new sweep, then press the following keys:
4MKR !5 MARKER ! HIGH
More 1 of 3 More 2 of 3 MARKER 1
More 3 of 3 NEXT PEAK
The two markers should be on the signals near the second and tenth vertical graticule lines (the rst graticule line is the
7.Record the
8. Press 4MKR5, More 1 of 3 , then MARKER ALL OFF on the EMC analyzer.
9.Change the equipment to the next settings listed in Table
10.On the EMC analyzer, press 4SGL SWP5. Wait for the completion of a new sweep, then press the following keys:
4MKR !5 MARKER ! HIGH
More 1 of 3 More 2 of 3 MARKER 1
More 3 of 3 NEXT PEAK
11.Record the
12.Repeat steps 8 through 11 for the remaining EMC analyzer span settings through TR Entry 6 (refer to Table
Performance Verication Tests