62.CISPR Pulse Response, HP 8590
92.On the EMC analyzer, press the following keys:
4DET5 QP ON OFF (ON)
93.Set the PERIOD to 50 ms on the pulse generator.
94.On the EMC analyzer, press the following keys:
4DET5 QP/AVG 10X OFF (10X)
4SGL SWP5
4MKR !5 MARKER ! HIGH
95.Record the marker amplitude reading in Table
96.Set PERIOD to 100 ms on the pulse generator. Press 4SGL SWP5
MARKER ! HIGH on the EMC analyzer.
Record the marker amplitude reading in Table
97.Set the PERIOD to 500 ms on the pulse generator. Press 4SGL SWP5 MARKER ! HIGH on the EMC analyzer.
Record the marker amplitude reading in Table
98.Set PERIOD to 980 ms on the pulse generator. Press 4SGL SWP5 MARKER ! HIGH on the EMC analyzer.
Record the marker amplitude reading in Table
99.Press 4TRIG5 on the pulse generator. Press 4SGL SWP5 on the EMC analyzer. Let the EMC analyzer sweep three divisions then press 4MAN5 on the pulse generator. Record the marker reading as the Isolated Pulse for Bands C and D in Table
100.Enter the Measured value for Band A 25 Hz Repetition Frequency as the Reference value for all the Repetition Frequencies listed for Band A (in Table