57.
HP 8593EM, HP 8594EM, HP 8595EM, HP 8596EM
Option 010
This performance verication test must be performed with the EMC analyzer set in the spectrum analyzer mode.
The tracking generator output is connected to the EMC analyzer input and the tracking is adjusted at 300 MHz for a maximum signal level. The tracking generator output is then connected to the input of a microwave spectrum analyzer. The tracking generator is tuned to several dierent frequencies, then the amplitude of the second and third harmonics relative to the fundamental are measured at each frequency.
For each output frequency, several sweeps are taken on the microwave spectrum analyzer over dierent frequency spans and the highest displayed spurious response is measured in each span. Responses at the fundamental frequency of the tracking generator output or its harmonics are ignored; they are tested in the \Harmonic Spurious Responses" performance verication test. The amplitude of the highest spurious response is recorded.
There are no related adjustments for this performance verication test.
Equipment Required
Spectrum analyzer, microwave
Cable, Type N, 62 cm (24 in)
Cable, BNC, 23 cm (9 in)
Adapter, Type N (m) to BNC (f)