39.Displayed Average Noise Level, HP 8591EM
16.Press 4MKR !5 MARKER ! HIGH and record the MKR frequency as the Measurement Frequency in Table
17.Press the following EMC analyzer keys:
4TRACE5 More 1 of 4 More 2 of 4 VID AVG ON OFF (OFF) DETECTOR SMP PK (SMP) (For Options 101, 102, and 301:
DETECTOR PK SP NG (SP))
4AUTO COUPLE5 IF BW AUTO MAN (AUTO) AVG BW AUTO MAN (AUTO)
4SPAN5 20 4kHz5
4FREQUENCY5
18.Set the center frequency to the Measurement Frequency recorded in Table
19.Press the following EMC analyzer keys:
4BW5 30 4Hz5
AVG BW AUTO MAN (MAN) 30 4Hz5
4SGL SWP5
Wait for the sweep to nish.
20. Press the following EMC analyzer keys:
4DISPLAY5 DSP LINE ON OFF (ON)
Adjust the display line so that it is centered on the average trace noise, ignoring any residual responses (refer to the Residual Responses verication test for any suspect residuals).
21.Record the display line amplitude setting as TR Entry 3 of the performance verication test record. The average noise level should be less than the specied limit.
Performance Verication Tests