25. Other Input Related Spurious Responses, HP 8593EM
Band 2
14.On the EMC analyzer, press 4FREQUENCY5, 9, 4GHz5.
15.Set the synthesized sweeper CW to 9 GHz.
16.Enter the power sensor 9 GHz CAL Factor into the measuring receiver.
17.Press the following EMC analyzer keys:
4MKR !5 MKR ! HIGH
4AMPLITUDE5 More 1 of 3 PRESEL PEAK
Wait for the CAL: PEAKING message to disappear, then press:
4MKR5 More 1 of 3 MARKER ALL OFF
18.Repeat steps 5 through 8 for the synthesized sweeper CW frequencies listed in Table
Band 3
19.On the EMC analyzer, press 4FREQUENCY5, 15, 4GHz5.
20.Set the synthesized sweeper CW to 15 GHz.
21.Enter the power sensor 15 GHz CAL Factor into the measuring receiver.
22.Press the following EMC analyzer keys:
4MKR !5 MKR ! HIGH
4AMPLITUDE5 More 1 of 3 PRESEL PEAK
Wait for the CAL: PEAKING message to disappear, then press:
4MKR5 More 1 of 3 MARKER ALL OFF
23.Repeat steps 5 through 8 for the synthesized sweeper CW frequencies listed in Table
Performance Verication Tests