25. Other Input Related Spurious Responses, HP 8593EM

Band 2

14.On the EMC analyzer, press 4FREQUENCY5, 9, 4GHz5.

15.Set the synthesized sweeper CW to 9 GHz.

16.Enter the power sensor 9 GHz CAL Factor into the measuring receiver.

17.Press the following EMC analyzer keys:

4MKR !5 MKR ! HIGH

4AMPLITUDE5 More 1 of 3 PRESEL PEAK

Wait for the CAL: PEAKING message to disappear, then press:

4MKR5 More 1 of 3 MARKER ALL OFF

18.Repeat steps 5 through 8 for the synthesized sweeper CW frequencies listed in Table 2-43 for Band 2.

Band 3

19.On the EMC analyzer, press 4FREQUENCY5, 15, 4GHz5.

20.Set the synthesized sweeper CW to 15 GHz.

21.Enter the power sensor 15 GHz CAL Factor into the measuring receiver.

22.Press the following EMC analyzer keys:

4MKR !5 MKR ! HIGH

4AMPLITUDE5 More 1 of 3 PRESEL PEAK

Wait for the CAL: PEAKING message to disappear, then press:

4MKR5 More 1 of 3 MARKER ALL OFF

23.Repeat steps 5 through 8 for the synthesized sweeper CW frequencies listed in Table 2-43 for Band 3.

Performance Veri￿cation Tests 2-161