54. Harmonic Spurious Outputs, HP 8591EM Option 010
REFERENCE LEVEL :: :: :: :: ::: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: : +5 dBm
RES BW : ::: :: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: : 30 kHz
LOG dB/DIV : :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: 10 dB
7.Set up the microwave spectrum analyzer by performing the following steps:
Note that the following steps are for an HP 8566A/B microwave spectrum analyzer, the steps may be dierent if you are using another microwave spectrum analyzer.
a.Press PEAK SEARCH and SIGNAL TRACK (ON). Wait for the signal to be displayed at center screen.
b.Press PEAK SEARCH, CF STEP SIZE 10 MHz, CENTER FREQUENCY, then SIGNAL TRACK (OFF).
c.Press CENTER FREQUENCY and the
d.Perform this step only if the Tracking Generator Output Frequency is less than 600 MHz. Press CENTER FREQUENCY and the
amplitude reading in Table
e. Press MARKER (OFF).
8.Change the microwave spectrum analyzer center frequency to the next frequency listed in Table
Table
Tracking Generator | 2nd Harmonic Level | 3rd Harmonic Level |
Frequency | (dBc) | (dBc) |
|
|
|
10MHz
100 MHz
300 MHz
850 MHz |
| N/A |
| ||
|
|
|
9.Locate the most positive 2nd Harmonic Level in Table