5. Frequency Readout and Marker Count Accuracy, HP 8593EM, HP 8595EM, and HP 8596EM

10.Set the synthesized sweeper CW to 1.5 GHz.

11.Set the EMC analyzer by pressing the following keys:

4FREQUENCY5 1.5 4GHz5

4BW5 IF BW AUTO MAN (MAN) 300 4Hz5

4SPAN5 20 4kHz5

12.Press 4MKR !5 MARKER ! HIGH on the EMC analyzer.

13.Record the MKR frequency reading as TR Entry 27 of the performance veri￿cation test record. The reading should be within the limits of 1.49999989 GHz and 1.50000011 GHz.

14.Set the EMC analyzer by pressing the following keys:

4BW5 IF BW AUTO MAN (MAN) 30 4Hz5

4SPAN5 2 4kHz5

15.Press 4MKR !5 MARKER ! HIGH 4MKR5 More 1 of 3 MK TRACK ON OFF

(ON), then wait until the count is completed (it may take several seconds).

16.Record the MKR reading as TR Entry 28 of the Performance veri￿cation Test Record. The reading should be within the limits of 1.49999989 and 1.50000011.

2-22 Performance Veri￿cation Tests