26. Other Input Related Spurious Responses, HP 8594EM

4MKR5 More 1 of 3 MK TRACK ON OFF (ON)

4SPAN5 200 4kHz5

Wait for the AUTO ZOOM message to disappear. Press the following analyzer keys:

4MKR !5 MKR ! HIGH More 1 of 3 MARKER !REF LVL

4MKR !5 MKR ! HIGH More 1 of 3 More 2 of 3 MARKER 1

4AMPLITUDE5 4 + 5 (step-down key) 4SGL SWP5

7.For each of the frequencies listed in Table 2-44 for a center frequency of 2.0 GHz, do the following:

a.Set the synthesized sweeper to the listed CW frequency.

b.Enter the appropriate power sensor Cal Factor into the measuring receiver.

c.Set the synthesized sweeper power level for a 010 dBm reading on the measuring receiver.

d.Press 4SGL SWP5 and wait for completion of a new sweep.

e.On the EMC analyzer, press 4MKR !5 MKR ! HIGH and record the MKR 1 amplitude reading in Table 2-44 as the Actual MKR 1 Amplitude.

The Actual MKR 1 Amplitude should be greater than the Minimum MKR 1 Amplitude listed in the table.

Note that the Minimum MKR 1 Amplitude is 10 dB more positive than the speci￿cation. This is due to the 10 dB change in reference level made in step 6.

Performance Veri￿cation Tests 2-167