26. Other Input Related Spurious Responses, HP 8594EM
4MKR5 More 1 of 3 MK TRACK ON OFF (ON)
4SPAN5 200 4kHz5
Wait for the AUTO ZOOM message to disappear. Press the following analyzer keys:
4MKR !5 MKR ! HIGH More 1 of 3 MARKER !REF LVL
4MKR !5 MKR ! HIGH More 1 of 3 More 2 of 3 MARKER 1
4AMPLITUDE5 4 + 5
7.For each of the frequencies listed in Table
a.Set the synthesized sweeper to the listed CW frequency.
b.Enter the appropriate power sensor Cal Factor into the measuring receiver.
c.Set the synthesized sweeper power level for a 010 dBm reading on the measuring receiver.
d.Press 4SGL SWP5 and wait for completion of a new sweep.
e.On the EMC analyzer, press 4MKR !5 MKR ! HIGH and record the MKR 1 amplitude reading in Table
The Actual MKR 1 Amplitude should be greater than the Minimum MKR 1 Amplitude listed in the table.
Note that the Minimum MKR 1 Amplitude is 10 dB more positive than the specication. This is due to the 10 dB change in reference level made in step 6.
Performance Verication Tests