57.
Figure
HP 8593EM, HP 8594EM, HP 8595EM, and HP 8596EM
Procedure
It is only necessary to perform step 1 if more than 2 hours have elapsed since a
The microwave spectrum analyzer should be allowed to warm up for at least 30 minutes before proceeding.
1.Perform a
Note that the following steps are for an HP 8566A/B microwave spectrum analyzer, the steps may be dierent if you are using another microwave spectrum analyzer.
a.Connect a BNC cable between CAL OUTPUT and RF INPUT.
b.Select the 2 0 22 GHz band, then press INSTR PRESET, 4RECALL5, 8. Adjust AMPTD CAL for a marker amplitude reading of 010 dBm.
c.Press 4RECALL5, 9. Adjust FREQ ZERO for a maximum amplitude response.
d.Press SHIFT, FREQUENCY SPAN to start the 30 second internal error correction routine.
e.When the CALIBRATING! message disappears, press SHIFT, START FREQ to use the error correction factors just calculated.
Performance Verication Tests