47. Fast Time Domain Sweeps, HP 8591EM Option 101 and Option 301

Table 2-61. Fast Sweep Time Accuracy

EMC Analyzer

Synthesizer/Function

Minimum

TR Entry

Sweep Time

Generator Frequency

Reading

(MKR 1)

 

 

 

 

18 ms

556 Hz

14.04 ms

1

10 ms

1 kHz

7.8 ms

2

1.0 ms

10 kHz

780 ￿s

3

100 ￿s

100 kHz

78 ￿s

4

20 ￿s

500 kHz

15.6 ￿s

5

 

 

 

 

Performance Veri￿cation Tests 2-317