50. Absolute Amplitude Accuracy, HP 8593EM, HP 8594EM, HP 8595EM, HP 8596EM Option 010
Figure
Absolute Amplitude Accuracy Test Setup,
HP 8593EM, HP 8594EM, HP 8595EM, and HP 8596EM
Procedure
1.Connect the Type N cable between the RF OUT 50 and INPUT 50 connectors on the EMC analyzer. See Figure
2.Press 4PRESET5 on the EMC analyzer, then wait for the preset routine to nish. Set the EMC analyzer by pressing the following keys:
4FREQUENCY5 300 4MHz5
4SPAN5 0 4Hz5
4BW5 IF BW AUTO MAN (MAN) 30 4kHz 5
4MKR5
4AUX/USER5 Track Gen SRC PWR ON OFF (ON) 102 4dBV5
3.Press TRACKING PEAK on the EMC analyzer, then wait for the PEAKING message to disappear.
4.Zero and calibrate the
5.Disconnect the Type N cable from the RF OUT 50 and connect the 100 kHz to 2.9 GHz power sensor to the RF OUT 50 . See Figure
6.On the EMC analyzer, press:
SRC PWR ON OFF (ON) 87 4dBV5
SRC PWR MAN AUTO (MAN) 91 4dBV5
4SGL SWP5
Performance Verication Tests