15. Reference Level Accuracy, HP 8593EM, HP 8594EM, HP 8595EM, and HP 8596EM

23.Set the synthesizer/level generator amplitude and EMC analyzer reference level according to Table 2-17.

24.At each setting, press 4SGL SWP5 on the EMC analyzer.

25.Record the MKR 1 amplitude reading in the performance veri￿cation test record as indicated in Table 2-17. The MKR 1 reading should be within the limits shown.

26.Repeat steps 23 through 25 for each entry in Table 2-17.

Table 2-17.

Reference Level Accuracy, Log Mode for Narrow

Bandwidths

Synthesizer/Level

EMC Analyzer

MKR 1 Reading (dB)

Generator Amplitude

Reference Level

 

 

 

 

 

 

 

 

(dBm)

(dB￿V)

Min.

TR Entry

Max.

 

 

 

 

 

010

87

0 (Ref)

0 (Ref)

0 (Ref)

0

 

 

 

 

97

00.4

19

+0.4

+10

107

00.5

20

+0.5

020

77

00.4

21

+0.4

030

67

00.5

22

+0.5

040

57

00.8

23

+0.8

050

47

01.1

24

+1.1

060

37

01.2

25

+1.2

070

27

01.3

26

+1.3

080

17

01.4

27

+1.4

 

 

 

 

 

27.Repeat steps 10 through 16 for the narrow bandwidths, using Table 2-18.

28.Record the MKR 1 amplitude reading in the performance veri￿cation test record as indicated in Table 2-18. The MKR 1 reading should be within the limits shown.

Performance Veri￿cation Tests 2-81