15. Reference Level Accuracy, HP 8593EM, HP 8594EM, HP 8595EM, and HP 8596EM
23.Set the synthesizer/level generator amplitude and EMC analyzer reference level according to Table
24.At each setting, press 4SGL SWP5 on the EMC analyzer.
25.Record the MKR 1 amplitude reading in the performance verication test record as indicated in Table
26.Repeat steps 23 through 25 for each entry in Table
Table
Reference Level Accuracy, Log Mode for Narrow
Bandwidths
Synthesizer/Level | EMC Analyzer | MKR 1 Reading (dB) | ||
Generator Amplitude | Reference Level |
|
|
|
|
|
|
|
|
(dBm) | (dBV) | Min. | TR Entry | Max. |
|
|
|
|
|
010 | 87 | 0 (Ref) | 0 (Ref) | 0 (Ref) |
0 |
|
|
|
|
97 | 00.4 | 19 | +0.4 | |
+10 | 107 | 00.5 | 20 | +0.5 |
020 | 77 | 00.4 | 21 | +0.4 |
030 | 67 | 00.5 | 22 | +0.5 |
040 | 57 | 00.8 | 23 | +0.8 |
050 | 47 | 01.1 | 24 | +1.1 |
060 | 37 | 01.2 | 25 | +1.2 |
070 | 27 | 01.3 | 26 | +1.3 |
080 | 17 | 01.4 | 27 | +1.4 |
|
|
|
|
|
27.Repeat steps 10 through 16 for the narrow bandwidths, using Table
28.Record the MKR 1 amplitude reading in the performance verication test record as indicated in Table
Performance Verication Tests