57.Non-Harmonic Spurious Outputs, HP 8593EM, HP 8594EM, HP 8595EM, HP 8596EM Option 010
or equal to 2000 MHz. Record this amplitude as the Highest Non-Harmonic Response Amplitude 2000 MHz as TR Entry 1 of the performance verication test record.
23.Locate in Table 2-75 the most-positive Non-Harmonic Response Amplitude for the microwave spectrum analyzer START frequency settings of greater than or equal to 2000 MHz. Record this amplitude as the Highest Non-Harmonic Response Amplitude 2000 MHz as TR Entry 2 of the performance verication test record.
Table 2-75. Non-Harmonic Responses Worksheet
Microwave Spectrum Analyzer | | Non-Harmonic Response |
| Settings | | | | Amplitude (dBc) | | |
| | | | | | | | | | | | | |
Start | Stop | Resolution | | | Center Frequency | | |
Frequency | Frequency | Bandwidth | | 9 kHz | | | | 1.5 GHz | | | 2.9 GHz |
| | | | | | | |
0.003 MHz* | 0.2 MHz | 3 kHz | | | | | | | | | | | |
| | | | | | | | | | |
0.2 MHz | 5.0 MHz | 30 kHz | | | | | | | | | | | |
| | | | | | | | | | |
5.0 MHz | 55 MHz | 100 kHz | | | | | | | | | | | |
| | | | | | | | | | |
55 MHz | 1240 MHz | 1 MHz | | | | | | | | | | | |
| | | | | | | | | | |
1240 MHz | 2000 MHz | 1 MHz | | | | | | | | | | | |
| | | | | | | | | | |
2000 MHz | 2900 MHz | 1 MHz | | | | | | | | | | | |
| | | | | | | | | | | | | |
* Adjust start frequency until the LO is just o the left side of the screen.
2-366 Performance Verication Tests