59.Tracking Generator Feedthrough, HP 8594EM Option 010

59.Tracking Generator F eedthrough, HP 8594EM Option 010

This performance veri￿cation test must be performed with the EMC analyzer set in the spectrum analyzer mode.

The tracking generator output is connected to the EMC analyzer input and the tracking is adjusted at 300 MHz for a maximum signal level. The tracking generator output is terminated and set for 108 dB￿V output power (maximum output power). The EMC analyzer input is also terminated. The noise level of the EMC analyzer is then measured at several frequencies.

There are no related adjustments for this performance veri￿cation test.

Equipment Required

Termination, 50 ￿ (two required)

Cable, Type N, 62 cm (24 in)

Cable, BNC, 23 cm (9 in)

Cable, Type N (m) to BNC (f)

Figure 2-84. Tracking Generator Feedthrough Test Setup, HP 8594EM