61.Tracking Generator LO Feedthrough Amplitude, HP 8593EM, HP 8594EM, HP 8595EM, HP 8596EM Option 010
SRC PWR ON OFF (ON) 01 4dBm5
4FREQUENCY5 9 4kHz5
4SGL SWP5
6.Connect the Type N cable from the tracking generator output to the microwave spectrum analyzer RF INPUT. See Figure
7.Set the microwave spectrum analyzer controls as follows:
CENTER FREQUENCY : :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: : 3.9217 GHz
SPAN ::: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: : 100 kHz
REFERENCE LEVEL ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: 0 dBm
RES BW : ::: :: :: :: :: ::: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: :: :: 1 kHz
LOG dB/DIV : : : : : : : : : : :: : : : : : : : :: :: : : : : : :: : : : :: : : : : : : : : : : : :: :: :: : : : : : :: 10 dB
8.On the microwave spectrum analyzer, press PEAK SEARCH and SIGNAL TRACK (ON), then wait for the signal to be displayed at center screen. Press
SIGNAL TRACK (OFF).
9.On the microwave spectrum analyzer, press PEAK SEARCH, PRESEL PEAK, then wait for the PEAKING! message to disappear.
10.Record the microwave spectrum analyzer marker amplitude in Table
11.Repeat steps 8 through 10 for the remaining EMC analyzer CENTER FREQ and microwave spectrum analyzer CENTER FREQUENCY settings listed in Table
12.Locate in Table
13.Locate in Table
Performance Verication Tests