27.Other Input Related Spurious Responses, HP 8595EM
5.Adjust the synthesized sweeper power level for a 010 dBm 60.1 dB reading on the measuring receiver.
6.On the EMC analyzer, press the following keys:
4MKR !5 MKR ! HIGH
4MKR5 More 1 of 3 MK TRACK ON OFF (ON)
4SPAN5 200 4kHz5
Wait for the AUTO ZOOM message to disappear. Press the following EMC analyzer keys:
4MKR !5 MKR ! HIGH
4MKR !5 More 1 of 3 MARKER !REF LVL
4MKR !5 MKR ! HIGH More 1 of 3 More 2 of 3 MARKER 1 4AMPLITUDE5 4+5
4SGL SWP5
7.For each of the frequencies listed in Table
a.Set the synthesized sweeper to the listed CW frequency.
b.Enter the appropriate power sensor CAL Factor into the measuring receiver.
c.Set the synthesized sweeper power level for 010 dBm reading on the measuring receiver.
d.Press 4SGL SWP5 and wait for the completion of a new sweep.
e.On the EMC analyzer, press 4MKR !5 MKR ! HIGH and record the
The Actual MKR 1 Amplitude should be greater than the Minimum MKR 1 Amplitude listed in Table
Note that the Minimum MKR 1 Amplitude is 10 dB more positive than the specication. This is due to the 10 dB change in reference level made in step 6.
Performance Verication Tests