24.Other Input Related Spurious Responses, HP 8591EM
24.Other Input Related Spurious Responses ,HP 8591EMThis performance verication test must be performed with the EMC analyzer set in the spectrum analyzer mode.
A synthesized source and the EMC analyzer are set to the same frequency and the amplitude of the source is set to 020 dBm (87 dBV). A
There are no related adjustment procedures for this performance test.
Equipment Required
Synthesized sweeper
Measuring receiver (used as a power meter)
Power sensor, 100 kHz to 1800 MHz
Adapter, Type N (f) to APC 3.5 (f)
Adapter, Type N (f) to Type N (f)
Cable, Type N, 183 cm (72 in)
Figure
Other Input Related Spurious Responses Test Setup, HP 8591EM
Performance Verication Tests