32. Spurious Response, HP 8595EM

Adapter, APC 3.5 (f) to APC 3.5 (f) (two required)

Adapter, Type N (m) to BNC (f)

Adapter, Type N (m) to APC 3.5 (f)

Adapter, Type N (f) to BNC (m)

Figure 2-48. Second Harmonic Distortion Test Setup, HP 8595EM

Procedure

This performance veri￿cation test consists of four parts:

Part 1: Second Harmonic Distortion, <2.9 GHz

Part 2: Second Harmonic Distortion, >2.9 GHz

Part 3: Third Order Intermodulation Distortion, <2.9 GHz

Part 4: Third Order Intermodulation Distortion, >2.9 GHz

Part 1: Second Harmonic Distortion, <2.9 GHz

1. Press 4PRESET5 on the synthesized sweeper, then set the controls as follows:

CW : ::: :: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: : 30 MHz

POWER LEVEL :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: ::: :: :: : 030 dBm

2.Connect the equipment as shown in Figure 2-48.

3.Press 4PRESET5 on the EMC analyzer, then wait for the preset routine to ￿nish. Set the EMC analyzer by pressing the following keys:

4FREQUENCY5 30 4MHz5

4SPAN5 1 4MHz5

4AMPLITUDE5 REF LVL 77 4+dB￿V5

4BW5 IF BW AUTO MAN (MAN) 30 4kHz5