This performance verication test must be performed with the EMC analyzer set in the spectrum analyzer mode.
The tracking generator output is connected to the EMC analyzer input and the tracking is adjusted at 300 MHz for a maximum signal level. The tracking generator output is then connected to the input of a microwave spectrum analyzer. The tracking generator is tuned to several dierent frequencies and the amplitude of the second and third harmonics relative to the fundamental are measured at each frequency.
There are no related adjustment procedures for this performance test.
Equipment Required
Spectrum analyzer, microwave
Cable, Type N, 62 cm (24 in)
Cable, BNC, 23 cm (9 in)
Adapter, Type N (m) to BNC (f)
Figure