18.Calibrator Amplitude Accuracy, HP 8590
HP 8590
This performance verication test must be performed with the EMC analyzer set in the spectrum analyzer mode.
This test measures the accuracy of the EMC analyzer CAL OUT signal. The rst part of the test characterizes the insertion loss of a Low Pass Filter (LPF) and 10 dB Attenuator. The harmonics of the CAL OUT signal are suppressed with the LPF before the amplitude accuracy is measured using a power meter.
Calibrator Frequency is not included in this procedure because it is a function of the Frequency Reference (CAL OUT Frequency = 300 MHz 6 [300 MHz 2 Frequency Reference]). Perform the Frequency Reference Accuracy test to verify the CAL OUT frequency.
Equipment Required
Synthesized sweeper
Measuring receiver (used as a power meter) Power meter
Power sensor, low power with a 50 MHz reference attenuator Power sensor, 100 kHz to 1800 MHz
Power splitter
10 dB Attenuator, Type N (m to f),
Cable, Type N, 152 cm (60 in) Adapter, APC 3.5 (f) to Type N (f)
Adapter, Type N (f) to BNC (m) (two required) Adapter, Type N (m) to BNC (f)
Procedure
This performance verication test consists of two parts:
Part 1: LPF, Attenuator and Adapter Insertion Loss Characterization
Part 2: Calibrator Amplitude Accuracy
Perform \Part 1: LPF, Attenuator and Adapter Insertion Loss Characterization"
before \Part 2: Calibrator Amplitude Accuracy."
A worksheet is provided at the end of this procedure for calculating the corrected insertion loss and the calibrator amplitude accuracy.
Performance Verication Tests