42. Displayed Average Noise Level, HP 8595EM
17. Press the following EMC analyzer keys:
4TRACE5
More 1 of 4 More 2 of 4 VID AVG ON OFF (OFF)
DETECTOR SMP PK (SMP) (For Options 101, 102, and 301:
DETECTOR PK SP NG (SP))
4AUTO COUPLE5 IF BW AUTO MAN (AUTO) AVG BW AUTO MAN (AUTO)
4SPAN5 10 4kHz5
4FREQUENCY5
Set CENTER FREQ to the Measurement Frequency recorded in Table
4BW5 IF BW AUTO MAN (MAN) 30 4Hz5
AVG BW AUTO MAN (MAN) 30 4Hz5
18.Press 4SGL SWP5 on the EMC analyzer, then wait for a new sweep to nish. Press the following EMC analyzer keys:
4DISPLAY5 DSP LINE ON OFF (ON)
Adjust the display line so that it is centered on the average noise trace, ignoring any residual responses (refer to Residual Response verication test for any suspected residuals).
Record the display line amplitude setting in the performance verication test record as indicated in Table
19. Press 4MKR5 and MARKER 1 ON OFF (OFF) to turn the marker o.