40. Displayed Average Noise Level, HP 8593EM

400 kHz

6. Press the following EMC analyzer keys:

4FREQUENCY5 400 4kHz5

4SPAN5 20 4kHz5

4AMPLITUDE5 37 dB￿V

4SWEEP/TRIG5 SWEEP CONT SGL (CONT)

7. Press the following EMC analyzer keys:

4BW5 30 4Hz5

4TRACE5 More 1 of 4 More 2 of 4

DETECTOR SMP PK (SMP) (For Options 101, 102, and 301:

DETECTOR PK SP NG (SP))

4SGL SWP5

Wait for the completion of a new sweep.

8. Press the following EMC analyzer keys:

4DISPLAY5 DSP LINE ON OFF (ON)

Adjust the display line so that it is centered on the average trace noise, ignoring any residual responses.

9.Record the display line amplitude setting as TR Entry 1 of the performance veri￿cation test record as the noise level at 400 kHz. The average noise level should be less than the speci￿ed limit.

1 MHz

10. Press the following EMC analyzer keys:

4FREQUENCY5 1 4MHz5

4SGL SWP5

Wait for the completion of a new sweep.

11. Press the following EMC analyzer keys:

4DISPLAY5 DSP LINE ON OFF (ON)

Adjust the display line so that it is centered on the average trace noise, ignoring any residual responses.

2-274 Performance Veri￿cation Tests