7. System Related Sidebands, HP 8590
RF :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: :: ON
2. Connect the equipment as shown in Figure
Option 026 only: Use the APC adapter to connect the cable to the EMC analyzer input.
3.Press 4PRESET5 on the EMC analyzer, then wait for the preset routine to
nish. Set the EMC analyzer to measure the system related sideband above the signal by pressing the following keys:
4FREQUENCY5 500 4MHz5
4SPAN5 10 4MHz5
4MKR !5 MKR ! HIGH
4MKR5 More 1 of 3 MK TRACK ON OFF (ON)
4SPAN5 200 4kHz5 4BW5 1 4kHz5
AVG BW AUTO MAN (MAN) 30 4Hz5
4.Allow the EMC analyzer to stabilize for approximately 1 minute, then press the following keys:
4MKR5 More 1 of 3 MK TRACK ON OFF (OFF)
4FREQUENCY5 CF STEP AUTO MAN (MAN) 130 4kHz5
4SGL SWP5
5.Wait for the completion of the sweep, then press the following EMC analyzer keys:
4MKR !5 MKR ! HIGH
More 1 of 3 More 2 of 3 MARKER 1 .
4FREQUENCY5
4 * 5
6.Measure the system related sideband above the signal by pressing 4SGL SWP5 on the EMC analyzer. Wait for the completion of a new sweep, then press
4MKR !5 MKR ! HIGH .
7.Record the
The system related sideband above the signal should be > 65 dBc.