46. Residual Responses, HP 8593EM, HP 8595EM, and HP 8596EM
Procedure
150 kHz to 1 MHz
1.Connect the termination to the EMC analyzer input as shown in Figure
2.Press 4PRESET5 on the EMC analyzer, then wait for the preset routine to nish. Press the following EMC analyzer keys:
4FREQUENCY5 More 1 of 2 Band Lock
4FREQUENCY5
START FREQ 150 4kHz5
STOP FREQ 1 4MHz5
4AMPLITUDE5 47 4dBV5 ATTEN AUTO MAN (MAN) 0 4dB5
4BW5 300 4Hz5
AVG BW AUTO MAN (MAN) 300 4Hz5
4DISPLAY5 DSP LINE ON OFF (ON) 17 4dBV5
3.Press 4SGL SWP5 and wait for a new sweep to nish. Look for any residual responses at or above the display line.
If a residual is suspected, press 4SGL SWP5 again. A residual response will persist on successive sweeps, but a noise peak will not. Note the frequency and amplitude of any residual responses above the display line and to the right of the marker in Table
Performance Verication Tests