34. Gain Compression, HP 8591EM

5.On the EMC analyzer, press 4PRESET5, then wait for the preset routine to ￿nish. Press the EMC analyzer keys as follows:

4FREQUENCY5 50 4MHz5

4SPAN5 20 4MHz5

4AMPLITUDE5 87 4dB￿V5

SCALE LOG LIN (LOG) 1 4dB5

4BW5 300 4kHz5

6.On the synthesized sweeper, adjust the power level for a 0 dBm reading on the measuring receiver. Set RF to o￿.

7.On the synthesizer/level generator, set the 50 ￿/75 ￿ switch to 50 ￿.

Note that the power level applied to the EMC analyzer input is 10 dB greater than the speci￿cation to account for the 10 dB attenuation setting. A power level of 0 dBm at the EMC analyzer input yields 010 dBm at the input mixer.

8.Disconnect the power sensor from the directional bridge and connect the directional bridge to the INPUT 50 ￿ connector of the EMC analyzer using an adapter. Do not use a cable.

9.On the EMC analyzer, press the following keys:

4MKR !5 MKR ! HIGH

4MKR5 More 1 of 3 MK TRACK ON OFF (ON)

4SPAN5 10 4MHz5

Wait for the AUTO ZOOM routine to ￿nish.

10.On the synthesizer/level generator, adjust the amplitude to place the signal 1 dB below the EMC analyzer reference level.

11.On the EMC analyzer, press 4MKR !5 MKR ! HIGH More 1 of 3

More 2 of 3 MARKER 1 .

12.On the synthesized sweeper, set RF to ON.

13.On the EMC analyzer, press 4MKR !5 MKR ! HIGH , then NEXT PEAK .

The active marker should be on the lower amplitude signal and not on the signal that is o￿ the top of the screen. If it is not on the lower amplitude signal, reposition the marker to this peak using the EMC analyzer knob.

2-238 Performance Veri￿cation Tests